Press the Snapshot button on the front panel to capture a 24-second period of numeric and waveform patient data. The unit captures 12 seconds proceeding and 12 seconds following the button press.
The X Series Advanced unit can store a minimum of 32 snapshots, including:
- Monitor snapshots
- Defibrillator snapshots
- Treatment snapshots
- Pacer snapshots
- Alarm snapshots
- Presenting Rhythm snapshots
Note: Presenting Rhythm snapshots are only taken at the start of a new patient. The snapshot will not be retaken if the unit is shut off for less than two minutes.
Reviewing and printing snapshots
You can perform this procedure from any main display screen.
- Press the Home/Display button repeatedly until the Trends window displays.
- Use the navigation keys to get to the Trends window.
- Press the Select button to scroll through the list of trends.
Note: The snapshots have a snapshot icon next to the time stamp.
- Select the desired snapshot from the Trends list and press Select. The snapshot numeric data displays.
- To print the data and the snapshot waveforms, highlight Print this Snapshot and press Select.
Comments
0 comments
Please sign in to leave a comment.